Digital 3/4" Combination pH Sensor, Convertible Sensor Style, PPS Body Material, General Purpose Glass pH Electrode
PPS, Convertible Body Style, 4.5m Analog Cable, General Purpose Glass pH Electrode, Pt 1000 Temperat...
View DetailsIntellical CDC401 is a digital, graphite, 4-pole conductivity cell with a temperature sensor. The laboratory version is available with a 3 or 1 m cable. The CDC401 is ideal for measuring electrical conductivity, salinity, resistivity, or total dissolved solids (TDS) in wastewater, drinking water, or general water quality applications.
Accuracy:Cond: ±0.5% of reading
Salinity: ±0.1, ±1 digit
TDS: ±0.5% ±1 digit
Cable Length:1 m
Diameter:15 mm
Dimensions (D x L):15 mm x 184 mm
Electrode Type:Conductivity Cell;
4 Poles - Graphite
Instrument:Sensor & Electrodes
Kit?:No
Length:184 mm
Method Type:Laboratory: 4 Poles - Graphite
Parameter:Conductivity, Resistivity, Salinity, TDS
Probe Type:Standard
Product Kit:Model: CDC401
Accessories Included: None
Range:Conductivity: 0.0 µS/cm - 200 mS/cm
TDS: 0.00 mg/L - 50.0 g/L as NaCl
Salinity: 0 - 42 ppt or ‰
Resistivity: 2.5 Ωcm - 49 MΩcm
Resolution:0.01/0.1 (5 digits max.)
Sample depth:45 mm
Sensor material:Noryl
Sensor Type:4-poles conductivity probe, Graphite, k = 0.40 cm-1
Temperature Accuracy:±0.3 °C
Temperature Range:"-10 - 110 °C"
Temperature Resolution:0.1 °C
Test requirements:Parameter Needed: Conductivity
Minimum Sample Depth (mm): 45
Thermistor: (ATC)
Stored time and date stamp for each measurement, operator and sample ID, calibration history, parameter, and probe serial number.
No guesswork needed to provide the most reliable and accurate results with each measurement.
Ideal for a multi-user environment with multiple HQD series Laboratory and Portable meters.