View DetailsThe low minimum detection limit and wide measurement range of the Hach 9184 sc Amperometric Free Chlorine Sensor makes it ideal for efficient residual chlorine monitoring. Continuously measures hypochlorous acid (HOCl).
Accuracy:2 % or ±10 ppb TFC, whichever is greater at pH <7.5
Accuracy 2:2 % or ±10 ppb HOCl, whichever is greater at pH <8
Calibration Interval:2 months
Calibration Method:Zero Calibration: electrically (automatic) or with de-chlorinated water
Calibration: Comparison of lab method with
process sample
pH Calibration: Single- or two-point using standards
or comparison of lab method with
process sample
Description :9184sc Hypochlorous Acid Chlorine Amperometric Sensor
Enclosure Rating:IP66 / NEMA 4X
Flow:0.25 L/h (200 - 250 mL/min) auto--regulated by flow through cell
Interferences:No interference from chloramines.
Chlorine dioxide and ozone interfere with
measurement.
Lower Limit of Detection (LOD):0.005 mg/L HOCl
Maintenance Interval:6 months Measurement Cell: for membrane and electrolyte, typical
Maintenance Interval 2:1 1.5 pH Cell: typical
Measuring Interval:Continuous
Measuring Principle:Amperometric/Membrane (electrode,
membrane, electrolyte)
Mounting:Flat, vertical surface (panel, stand, etc.)
Operating Temperature Range:0 - 45 °C
Parameter:Hypochlorous Acid Only Chlorine
pH Range:4 - 8, acidification unit available for pH > 8
Pressure Range:0.1 - 2 bar in flow cell
Range:0 - 20 mg/L hypochlorous acid HOCl
Relative Humidity:0 - 90 % non-condensing
Response time:90 % in less than 90 seconds
Sample Input:0.25 " outer diameter
Sample Output:0.5 " inner diameter
Sample Temperature:2 - 45 °C
Standard deviation:0.7 %
Storage conditions:-20 - 60 °C
Temperature Compensation:Automatic over sample temperature range
Automatic over sample temperature range
The Hach 9184sc sensor has a minimum detection limit of 5 ppb or 0.005 mg/L hypochlorous acid in the form of HOCl.
The 9184sc sensor is particularly useful for drinking water distribution monitoring, chlorination applications, demineralizer systems, wastewater, or cooling water processes due to its 0 - 20 ppm measurement range.
Measurements are made continuously by the sensor and reported to the controller
The 9184sc Analyzer is designed to be used with all Hach sc digital controllers. Set up is fast and easy, and the system is compatible with the full line of Hach digital sensors.