SC200 Controller, 100-240 VAC, one digital sensor input, one analog pH/ORP/DO sensor input, HART, 4-20 mA outputs
The SC200 Universal Controller is the most versatile controller on the market. It allows the use of...
View DetailsThe 5500 sc silica analyzer needs only two liters of reagent to perform unattended for up to 90 days; twice as long as the Series 5000. The industry's only pressurized reagent delivery system eliminates the frequent maintenance associated with pumps. Prognosys which is a predictive diagnostic system allows you to be proactive in your maintenance. Warning LEDs, and high-visibility notification screens let you avoid unplanned downtime. No more dripping reagents on the instrument, the floor, or your clothing while fumbling with tubes and straws. Simply match the color-coded cap to the sealed reagent bottle and twist gently. Grab Sample In and Grab Sample Out features allow quick analysis of a grab sample poured into the analyzer, and facilitate taking a sample out of the online silica analyzer to verify in a lab test.
Accuracy:0 - 500 µg/L: ±1% or ±1 µg/L of reading, whichever is greater; 500 - 5000 µg/L: ±5%
Alarm:Four relays output; type: not powered SPDT relays, each rated at
5 A resistive, 240 VAC maximum
Connection: 18 to 14 AWG wire, 18 AWG stranded recommended
Application:Pure water / Power
Certifications:CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 )
KC (EN 61326-1: 2006)
C-tick (EN 61326-1: 2006)
cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012)
Controller Compatibility:SC200, SC1000
Enclosure Rating:IP56 / NEMA 4X
Fitting type:Sample line and sample bypass drain: 6 mm (¼-in.)
Air purge inlet: 6 mm (¼-in.)
Grab Sample:Grab Sample In and Grab Sample Out capability
Light source:Class 1M LED (light emitting diode) with a peak
wavelength of 810 nm
Lower Limit of Detection (LOD):0.5 µg/L
Measuring Principle:Colorimetric
Mounting:Wall, panel or table
Number of Channels:1
Operating Humidity:5 - 95 % non-condensing (indoor use only)
Operating Temperature Range:5 - 45 °C
Outputs:4 - 20 mA (4x)
Parameter:Silica
Pollution Degree:2/II
Power:100-240 VAC, 50/60 Hz; 0.5 A nominal, 8.3 A maximum
Accessory output: 100-240 VAC, 5.0 A maximum
Power requirements (Hz):50/60 Hz
Power requirements (Voltage):100 - 240 VAC
Protection Class:I
Range:0 - 5000 µg/L as SiO2
Reagent Consumption:2 L of each reagent every 90 days with 15 minute cycle time
Relays:Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Repeatability:± 0.5 µg/L or ± 1 % of reading, whichever is greater
Response time:Typically, 9.5 minutes at 25 °C (77 °F); changes with temperature
Sample Flow Rate:55 - 300 mL/min
Sample Pressure:2 - 87 psi (to Preset Pressure Regulator)
Sample Temperature:5 - 50 °C (41 - 122°F)
Storage conditions:-20 - 60 °C